Page 31
        
        
          Jenway Catalogue
        
        
          Fluorimeters
        
        
          Technical Specification
        
        
          
            Parameter
          
        
        
          Wavelength selection
        
        
          Intelligent Filter Modules (IFM)
        
        
          Light source
        
        
          Pulsed Xenon lamp with press-to-read operation
        
        
          Expected life of lamp
        
        
          No replacement required in normal operation
        
        
          Raw fluorescence readings
        
        
          Yes
        
        
          Concentration standard curve
        
        
          Blank and up to 6 standards
        
        
          Kinetics, timed option
        
        
          In both raw fluorescence and concentration modes
        
        
          Kinetics options
        
        
          1 reading every 3 to 999 seconds
        
        
          No. methods stored
        
        
          20
        
        
          No. results stored
        
        
          100 per method
        
        
          GLP support
        
        
          Method lock with password access
        
        
          Time and date stamped results
        
        
          Last calibration date and time stored
        
        
          Calibration interval set and Calibration due reminder
        
        
          User ID
        
        
          Outputs
        
        
          RS232
        
        
          PC software
        
        
          Compatible with DataWay (option)
        
        
          Sample handling
        
        
          10mm square cuvette holder (standard)
        
        
          Heated sample holder (option)
        
        
          Sipper pump (option)
        
        
          Power supply
        
        
          90 to 230V universal mains adapter
        
        
          Size (w x d x h), mm
        
        
          365 x 272 x 160
        
        
          Weight, kg
        
        
          6
        
        
          
            6270
          
        
        
          
            6280
          
        
        
          
            6285
          
        
        
          Wavelength range
        
        
          190 to 1100nm
        
        
          190 to 650nm 190 to 850nm
        
        
          Sensitivity*
        
        
          1µg/ml
        
        
          <1ng/ml
        
        
          <1ng/ml
        
        
          Dynamic range*
        
        
          5.2x10
        
        
          4
        
        
          5.2x10
        
        
          6
        
        
          5.2x10
        
        
          6
        
        
          Detector
        
        
          Photodiode
        
        
          Photomultiplier tube Photomultiplier tube
        
        
          * Using quinine sulphate
        
        
          62 Series
        
        
          Note: Excitation and emission filters are not included and should be purchased as separate items. Please refer to the
        
        
          filter section.
        
        
          
            Part Code Description
          
        
        
          627 001
        
        
          6270 fluorimeter supplied with 100 disposable cuvettes (060 247) and mains adapter for use on 90 to 230V
        
        
          supply with UK, Euro and US mains leads
        
        
          628 001 6280 fluorimeter supplied with 100 disposable cuvettes (060 247) and mains adapter for use on 90 to 230V
        
        
          supply with UK, Euro and US mains leads
        
        
          628 501 6285 fluorimeter supplied with 100 disposable cuvettes and (060 247) mains adapter for use on 90 to 230V
        
        
          supply with UK, Euro and US mains leads
        
        
          Ordering Information
        
        