Page 27
Jenway Catalogue
PFP7/C
Part code: 500 801
Clinical Flame Photometer
The PFP7/C is a lowtemperature, single channel flame photometer
that is specifically designed for use in clinical applications for the
determination of sodium, potassium and lithium. The in-built
lineariser circuitry of the PFP7/C enables readings of both sodium
and potassium to be displayed directly in mmol/l.
Technical Specification
Range
120 to 160mmol/l Na
0 to 10.0mmol/l K
Limits of detection Na 0.2ppm
K 0.2ppm
Li 0.25ppm
Reproducibility
<1% coefficient of variation for 20
consecutive samples using 10ppm Na
set to read 50
Linearity
<2% error when 3ppm Na/K and
5ppm Li are set to read 100
Stability
<2% drift over 5 min when
continuously aspirating 10ppm sample
set to 50.0
Specificity
Interference from Na/K and Li equal in
concentration to test element will be
<0.5%
Recorder output
Nominal 1.00V for a reading of 100.0
Services
Electrical
90-125V or 190-250V @50/60Hz*
Air
Moisture and oil free, 6 litre/min
at 14psi
Fuel
Propane, butane, natural gas or LPG
Size (w x d x h), mm 420 x 360 x 300
Weight, kg
8
PFP7/C
• Designed for clinical analysis
• Supplied with Na, K and Li filters
• Low temperature, single channel
• Flame failure safety system
• Operates with propane, butane, natural gas or
LPG
Key Features
Part Code Description
500 801 PFP7/C clinical flame photometer supplied with
Na, K and Li filters, connecting hoses and clips,
compressor plug and drain trap (230V/50Hz)
Ordering Information
Note: This product will only operate with the correct type of
compressor and gas regulator as specified.
* Voltage variants available see page 94