Page 13
Jenway Catalogue
4520
Part code: 452 001
• Conductivity, resistivity, TDS and salinity modes
• Special pure water mode for ultra low
conductivity measurements
• Storage of 500 results with built in data logger
• Infra-red data link (IrDA) for connection to
printer and RS232 link to connect to printer or
PC via DataWay software
Bench Conductivity/TDS Meter
The 4520 is a high specification laboratory conductivity /resistivity/
TDS/salinity and temperature meter that offers additional
accuracy with 1, 2 or 3 point conductivity calibration across the
wide measurement range of 0 to 19.99S. The dedicated ‘pure
water’ mode ensures the optimum accuracy for this difficult
application.
Technical Specification
Conductivity
Range
6 auto-selected from 0 to 19.99S*
Resolution
0.01µS -0.01S
Accuracy
±0.5% ±2 digits
Automatic std. recognition 10µS, 84µS, 1413µS, 12.88mS
Cell constant
0.010 to 19.999
TDS
Range
6 auto selected ranges 0-1999g/l*
Resolution
0.01mg/l -1g/l
Accuracy
±0.5% ±2 digits
EC ratio
0.50 to 0.80
Resistivity
Range
0-20Mohm
Resolution
0.01Mohm
Accuracy
±0.5% ±2 digits
Salinity
Range
0-99.9g/l
Resolution
0.1g/l
Accuracy
0-35 ±1g/l, 35-99.9 ±3g/l
Temperature
Range
-10 to 105°C
Resolution
0.1°C
Accuracy
±0.5°C
ATC and manual
0 to 100°C
Temperature coefficient
0.00% to 4.00%/ºC
Reference temperature 18, 20 or 25°C
GLP
Calibration reminder interval,
alarm outputs and security code
protected data
Outputs
Analogue, RS232 and IrDA interface
Connector
7-pin DIN
Power
9V AC ±10% @ 50/60Hz**
Size (l x w x d), mm 210 x 250 x 55
Weight, g
850
*With K=10 cell
** Voltage variants available see page 94
Key Features
4520
Ordering Information
Part Code Description
452 001 4520 conductivity meter supplied with glass
conductivity probe with ATC (K=1) (027 013),
electrode holder (903 300) and UK power supply
(021 030)
1...,3,4,5,6,7,8,9,10,11,12 14,15,16,17,18,19,20,21,22,23,...104