Page 12
Jenway Catalogue
4510
Part code: 451 001
Bench Conductivity/TDS Meter
4510
• Simultaneous display of conductivity or
TDS and temperature
• Automatic selection of range for best
resolution
• Calibration by cell constant or standard
solutions
• Automatic standard recognition
• Storage of up to 32 results
• Bi-directional RS232 link to printer or PC
Key Features
Part Code Description
451 001 4510 conductivity meter supplied with glass
conductivity probe with ATC (K=1) (027 013),
electrode holder (903 300) and UK power supply
(021 030)
Ordering Information
The 4510 is easy to use but with the flexibility to meet the
broadest range of applications and for those where greater
accuracy is required the 4510 has automatic conductivity standard
recognition and endpoint detection. Set-up options include cell
constant, temperature coefficient and reference temperature.
The instrument can store up to 32 readings, which can also be
sent to a printer or transferred to computer via DataWay and
the RS232 interface.
Technical Specification
Conductivity
Range
6 auto-selected
from 0 to 19999mS*
Resolution
0.01µS to 1mS
Accuracy
±0.5% ±2 digits
Automatic std. recognition 10µS, 84µS, 1413µS, 12.88mS
Cell constant
0.010 to 19.999
TDS
Range
6 auto-select from 0-19999g/l*
Resolution
0.01mg/l to 1g/l
Accuracy
±0.5% ±2 digits
EC Ratio
0.50 to 0.80
Temperature
Range
-10 to 105°C
Resolution
0.1°C
Accuracy
±0.5°C
ATC and manual
0 to 100°C
Temperature coefficient
0.00% to 4.00%/ºC
Reference temperature
18, 20 or 25°C
Outputs
Analogue and RS232
Connector
7-pin DIN
Power
9V AC ±10% @ 50/60Hz**
Size (l x w x d), mm
210 x 250 x 55
Weight, g
850
*With K=10 cell
** Voltage variants available see page 94