Page 75
        
        
          Jenway Catalogue
        
        
          Visible and UV/Visible Scanning Spectrophotometers
        
        
          Technical Specification
        
        
          
            Wavelength
          
        
        
          Range
        
        
          320 to 1100nm (6700)
        
        
          190 to 1100nm (6705, 6715)
        
        
          Spectral bandwidth 4nm (6700, 6705)
        
        
          1.5nm (6715)
        
        
          Resolution
        
        
          0.1nm
        
        
          Accuracy
        
        
          ±1.0nm
        
        
          Repeatability
        
        
          ±0.2nm
        
        
          
            Photometrics
          
        
        
          Transmittance
        
        
          0 to 199.9%T
        
        
          Absorbance
        
        
          -0.300 to 3.000A
        
        
          Accuracy
        
        
          ±0.005A at 1.0A
        
        
          Resolution
        
        
          0.1%T, 0.001A
        
        
          Stability
        
        
          <0.001A/h after warm-up
        
        
          Stray light
        
        
          <0.1%T at 340nm (6700) <0.05% at 220nm (6705, 6715)
        
        
          
            Quantitation
          
        
        
          Range
        
        
          -99999 to +99999
        
        
          Calibration
        
        
          Up to 20 standards with 5 replicates of each
        
        
          Units
        
        
          mEq/l, ppm, mg/l, g/l, %, µg/ml, mg/ml, g/dl, mg/dl, µg/l, ng/l, µg/dl, M, mM, µM, U/l, mU/l,
        
        
          U/ml, blank
        
        
          Curve fit algorithms
        
        
          Linear, quadratic and cubic functions
        
        
          
            Multi-wavelength
          
        
        
          Data points
        
        
          Up to 4 wavelengths
        
        
          Calculations
        
        
          Ratio, difference, formulae with factors
        
        
          
            Kinetics
          
        
        
          Time limits
        
        
          0 to 9999 seconds
        
        
          Calibration
        
        
          Standard or factor
        
        
          Analysis
        
        
          Curve details with mean rate of change plus formula of line of best fit
        
        
          
            Spectrum
          
        
        
          Range
        
        
          Any range between 320 (6700) or 190 (6705, 6715) and 1100nm
        
        
          Scan speed
        
        
          1500nm/min
        
        
          Scan interval
        
        
          0.1nm
        
        
          Analysis
        
        
          Auto peaks and valleys, zoom, addition, subtraction, peak ratios, smoothing, area under
        
        
          curve, wavelength table, derivatives, analysis points, overlay
        
        
          
            Other
          
        
        
          Light source
        
        
          Tungsten halogen lamp (6700)
        
        
          Xenon lamp (6705, 6715)
        
        
          Configuration
        
        
          Secure multi-user and free access
        
        
          Number of users
        
        
          10 users + supervisor
        
        
          Methods/results memory >500 on internal flash memory or removable media
        
        
          Removable media
        
        
          SD memory card
        
        
          Outputs
        
        
          USB, Centronics, Analogue, Internal printer
        
        
          Size (w x d x h), mm 490 x 390 x 220
        
        
          Weight, kg
        
        
          7.5
        
        
          NOTE: All 67 Series spectrophotometers are supplied with 2GB of internal memory, 2GB SD memory card, 100 disposable cuvettes, instruction manual,
        
        
          power cable and PC software on CD ROM with interface cable.
        
        
          67 Series
        
        
          
            Part Code Description
          
        
        
          670 0B0
        
        
          6700 Visible Scanning Spectrophotometer fitted with a single 10 x 10mm cuvette holder
        
        
          670 5B0 6705 UV/Visible Scanning Spectrophotometer fitted with a single 10 x 10mm cuvette holder
        
        
          671 5B0 6715 UV/Visible Scanning Spectrophotometer fitted with a single 10 x 10mm cuvette holder
        
        
          Ordering Information
        
        
          Jenway Catalogue