Page 44
        
        
          Jenway Catalogue
        
        
          370
        
        
          Part code: 370 271
        
        
          
            Part Code Description
          
        
        
          370 271 370 pH/mV meter supplied with epoxy  bodied
        
        
          combination electrode (924 001), ATC probe
        
        
          (027 500), pH buffer sachets, batteries (fitted)
        
        
          and free carry case
        
        
          Ordering Information
        
        
          • pH, absolute and relative mV ranges
        
        
          • Automatic or manual calibration
        
        
          • Automatic or manual temperature
        
        
          compensation
        
        
          • Simultaneous pH and temperature reading
        
        
          • Storage of up to 32 results
        
        
          Enterprise Portable pH/mV Meter
        
        
          The 370 hand held pH/mV meter offers simple, yet professional
        
        
          performance with 1 or 2 point calibration and automatic buffer
        
        
          recognition with manual override. The unit simultaneously
        
        
          displays temperature compensated pH readings or electrode
        
        
          potential and temperature.
        
        
          A comprehensive range of application specific electrodes and
        
        
          consumables are available for use with the 370 meter.
        
        
          Technical Specification
        
        
          
            pH
          
        
        
          Range
        
        
          -2.00 to 16.00
        
        
          Resolution
        
        
          0.01
        
        
          Accuracy
        
        
          ±0.02
        
        
          Automatic buffer recognition 4.00, 7.00, 9.22, 10.00 (with
        
        
          manual override)
        
        
          Calibration
        
        
          User selectable 1 or 2 point
        
        
          
            mV (absolute or relative)
          
        
        
          Range
        
        
          ±1999mV
        
        
          Resolution
        
        
          1mV
        
        
          Accuracy
        
        
          ±1mV
        
        
          
            Temperature
          
        
        
          Range
        
        
          -10 to +105°C
        
        
          Resolution
        
        
          0.1°C
        
        
          Accuracy
        
        
          ±0.5°C
        
        
          ATC/Manual range
        
        
          0 to 100°C
        
        
          Connector
        
        
          BNC
        
        
          Power
        
        
          2 AA batteries
        
        
          Battery life
        
        
          500 hours typical
        
        
          Size (l x w x d), mm
        
        
          175 x 75 x 35
        
        
          Weight, g
        
        
          250
        
        
          Key Features
        
        
          370